Patient Monitoring

Patient Monitoring

Showing all 4 results

  • ,

    368,01 лв.

    Riester ri-fox N Pulse Oximeter

    • Affordable mobile precision
    • For the fast and precise measurement of pulse and SpO2 for adults and children in just seconds.

    Product details

    • Small, mobile device with integrated sensor
    • Simple and fast measurement by attaching to the finger
    • Bright LED illumination for good visibility
    • Battery lifetime of approx. 30 hours (in continuous operation)
    • Automatic switch-off after 8 seconds
    • 30 months warranty

    Technical specifications

    • SpO2 measurement range 70%-100%
    • SpO2 resolution 1%
    • SpO2 measurement accuracy 70%-100%: +/- 2%
    • Pulse rate measurement range 30-235bpm
    • Pulse rate resolution 1bpm
    • Pulse rate measurement accuracy 30-100: +/-2bpm, 101-235: +/-2%
    • Display LED (SpO2, pulse rate, pulse signal, visual indicator for low battery)
    • Environmental Operating temperature 5°C – 40°C (41°F – 104°F)
    • Storage temperature -20°C – 55°C (-4°F – 131°F)
    • Operating humidity <= 80%, no condensation
    • Storage humidity <=93%, no condensation
    • Energy supply 2 batteries AAA, alkaline type LR03 (included)
    • Dimensions 66 x 39 x 32 mm (2.60 x 1.54 x 1.26 in)
    • Weight 48 g (1.59 oz) with batteries
    • Disinfection: Surface disinfection

    Specifications

    Height3 cm
    Weight50 gm
    Breadth5 cm
    Length8 cm

     

    Select options
  • ,

    From 4801,04 лв.

    Riester RVS 100 Advanced Vital Signs Monitor

    Advanced vital signs monitor for pulse oximetry, blood pressure and temperature measurements.

    Riester’s RVS-100 is an advanced vital signs monitor which offers pulse oximetry, blood pressure and temperature measurements. The device has ‘Monitor‘, ‘Spot-Check‘ and ‘Triage‘ profile options and a high performance touchscreen.

    The RVS – 100 can communicate through a wired or wireless connection to the hospital’s EMR, according to the HL7 standard. The highly flexible, modular design offers many configurations and settings to suit the needs of hospitals, clinics and private practices, as well as long-term care settings.

    • 3 measurement modes (Monitor, Spot-check, Triage)
    • High-sensitivity 8” TFT colour touch screen with self-explanatory user interface and integrated help-     function
    • Displays SYS, DIA, MAP, Pulse, Temp, Plethysmogram
    • Clinical grade NIBP* including averaging mode
    • Programmable NIBP averaging mode (2-5 readings)
    • Riester, Masimo* or Nellcor* Oximax SpO2
    • Predictive Thermometer (Filac 3000)
    • Connectivity with the EMR – follows the HL7 standard
    • LAN Connectivity as standard
    • Programmable alarms (audible, visible) with 3 priority levels
    • Supports nurse call RJ11 port
    • Available as wall, desk and floor model
    • Can be expanded to wall diagnostic station
    • Internal memory for 5,000 measurements
    • Lithium-ion battery with up to 17 hours ON time

    Options

    • Thermal printer (internal)
    • Barcode reader
    • Mobile stand including basket for storage (cuffs, sensors, cables)
    • WiFi

    Accessories supplied with monitor

    • NIBP extension tube
    • 2 cuffs (adult & adult large incl. connectors)
    • SpO2 sensor adult
    • SpO2 extension cable
    • Oral temperature probe
    • Probe cover

    Technical Data

    Monitor

    • Power Supply: 100 – 240 V AC, 50/60 Hz
    • Battery: Lithium-Ion battery, rechargeable, 10.8 V DC, 6,600 mAh,

    Charging time to a 100%: 6 hours,

    • Operating time approx. 11 hours (new and fully charged battery at 25°C ambient temperature with connected SpO2, Temp, and NIBP on AUTO mode (15 minutes interval))
    • Low battery indicator
    • Charging battery indicator
    • Dimensions Monitor: L x H x W: 303 x 230 x 110 mm (11.93 x 9.06 x 4.33 in)
    • Weight: 3.60 kg (7.94 lb) including temp module, printer
    • Dimensions Basket: L x H x W: 270 x 225 x 200 mm (10.63 x 8.86 x 7.88 in)

    Alarm:

    • Visible LEDs (Yellow, Red, Blue, Green, Orange)
    • Audible Speaker
    • Gives audible alarm, QRS tone
    • Supports Pitch Tone and multi-level volume
    • Alarm tones meet the requirement of IEC 60601-1-8.
    • Alarm Pressure: 45 dB to 85 dB test distance is 1 meter from the tone.

    Ports: 

    • 1 x AC power inlet
    • 1 x standard RJ45 interfaces.100 BASE-TX, IEEE 802.3
    • 4 x USB ports
    • 1 x Equipotential grounding point
    • 1 x RJ11 connector for nurse call
    • DC out port 15V/1.2 A   Clinical-grade NIBP (non-invasive blood pressure)

    Clinical-grade NIBP (non-invasive blood pressure)

    • Measuring method: automatic oscillometric
    • Measurement range:
      • Adult SYS 30 – 270 mmHg, DIA 10 – 220 mmHg, MAP 20 – 235 mmHg
      • Pediatric SYS 30 – 235 mmHg, DIA 10 – 220 mmHg, MAP 20 – 225 mmHg
      • Neonate SYS 30-135 mmHg, DIA 10-110 mmHg, MAP 20-125 mmHg
    • Cuff pressure range: 0 – 280 mmHg
    • Resolution: 1 mmHg
    • Pressure accuracy:
      • Static ±3 mmHg
      • Clinic ±5 mmHg, standard deviation ≤ 8 mmHg
    • Pulse rate range: 40 – 240 bpm
    • Pulse rate accuracy: 1 bpm
    • Initial inflation pressure:
      • Adult 160 mmHg, Pediatric 130 mmHg, Neonate 75 mmHg
      • Double overpressure protection (hardware and software)
      • Adult 297 ± 3 mmHg, Pediatric 252 ± 3 mmHg , Neonate 147 ± 3 mmHg
    • Alarm range:
      • SYS : 0 – 300 mmHg
      • DIA : 0 – 300 mmHg
      • MAP : 0 – 300 mmHg

    Riester SpO2 

    • Measurement range: 0 – 100%
    • Resolution: 1%
    • Accuracy:
      • ±2% (70 to 100%)
      • ±3% (40% to 69%)
      • Unspecified (0% to 39%)
    • Alarm range: 0 – 100%
    • Average time: 4 sec, 8 sec, 16 sec
    • Pulse rate
      • Measurement range: 20 – 250 bpm
      • Resolution: 1 bpm
      • Accuracy: ±1% or ±1 bpm whichever is greater

    Nellcor SpO2 

    • Measurement range: 0-100%
    • Resolution: 1%
    • Accuracy:
      • 70 – 100%: ±2% (adult, pediatric)
      • 70 – 100%: ±3% (neonate)
      • Unspecified (0 – 69%)
    • Alarm range: 0 – 100%
    • Average time: 8 sec, 16 sec
    • Pulse rate
      • Measurement range: 20 – 300 bpm
      • Resolution: 1 bpm
      • Accuracy: 20 – 250 bpm: ±3 bpm & 251 – 300 bpm: unspecified

    Masimo SpO2

    • Measurement range: 0 – 100%
    • Resolution: 1%
    • Accuracy:
      • 70 – 100%: ±2% (adult, pediatric, non-motion conditions)
      • 70 – 100%: ±3% (neonate, non-motion conditions)
      • 70 – 100%: ±3% (motion conditions)
      • 0 – 69%: unspecified
    • Average time: 2 – 4 sec, 4 – 6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec
    • Pulse rate
      • Measurement range: 25 – 240 bpm
      • Resolution: 1 bpm
      • Accuracy: ±3 bpm (non-motion conditions) & ±5 bpm (motion conditions)

    Predictive thermometer

    • Oral ( Quick Mode )
      • non-fever temps: 3-5 seconds
      • fever temps: 8-10 seconds
    • Oral ( Standard Mode ): 6-10 seconds
    • Axillary Mode: 8-12 seconds
    • Rectal Mode: 10-14 seconds
    • Direct Mode (all sites): 60-120 seconds
    • Measurement accuracy: 35.5ºC – 42ºC (95.9°F – 107.6°F)
    • Direct Mode (all sites): ±0.1ºC (±0.2ºF)
    • Standard Prediction Mode (all sites): ±0.1ºC (±0.2ºF)
    • Quick Prediction Mode (oral only): ±0.3ºC (±0.5ºF)
    Select options
  • ,

    From 5934,07 лв.

    Riester RVS 200 Integrated Modular Wall Diagnostic Station

    Unique modular combination of an advanced vital signs monitor and EENT diagnostic instruments.

    Riester’s RVS-200 is the unique modular combination of an advanced vital signs monitor (RVS-100) and ri-scope® L EENT diagnostic instruments with HighPerformance LED illumination.

    The RVS – 200 Wall Diagnostic System can communicate through a wired or wireless connection to the hospital’s EMR, according to the HL7 standard. The highly flexible, modular design offers many configurations and settings to suit the needs of hospitals, clinics as well as primary
    care and long-term care settings.Product detailsDiagnostic handles
    • ri-scope® L instruments with renowned optics and advanced LED lighting
    • Maintenance-free LED technology (4,200 K, CRI >90)

    Monitor

    • 3 measurement modes (Monitor, Spot-check, Triage)
    • High-sensitivity 8” TFT colour touch screen with self-explanatory user interface and integrated help-function
    • Displays SYS, DIA, MAP, Pulse, Temp, Plethysmogram
    • Clinical grade NIBP* including averaging mode
    • Programmable NIBP averaging mode (2-5 readings)
    • Riester, Masimo* or Nellcor* Oximax SpO2
    • Predictive Thermometer (Filac 3000)
    • Connectivity with the EMR – follows the HL7 standard
    • LAN Connectivity as standard
    • Programmable alarms (audible, visible) with 3 priority levels
    • Supports nurse call RJ11 port
    • Internal memory for 5,000 measurements

     (*includes neonate mode)

    Options

    • Thermal printer (internal)
    • Barcode reader
    • Extension modul (extension of the entire system up to 3 handles)
    • WiFi

    Accessories supplied with monitor

    • NIBP extension tube
    • 2 cuffs (adult & adult large incl. connectors)
    • SpO2 sensor adult
    • SpO2 extension cable
    • Oral temperature probe
    • Probe cover

    Technical Data

    Power Supply: 100 – 240 V AC, 50/60 Hz
    Dimensions: L x H x W: 759 x 310 x 110 mm (29.88 x 12.21 x 4.33 in)
    Weight: 7.60 kg (16.76 lb) including temp module, printer
    Dimensions Basket: L x H x W: 270 x 225 x 200 mm (10.63 x 8.86 x 7.88 in)
    Specifications of the monitor please see RVS-100.
    Select options
  • ,

    347,47 лв.

    SpO2 Sensors for RVS 100 Advanced Vital Signs Monitor

    Riester SpO2

    • Measurement range: 0 – 100%
    • Resolution: 1%
    • Accuracy:
      • ±2% (70 to 100%)
      • ±3% (40% to 69%)
      • Unspecified (0% to 39%)
    • Alarm range: 0 – 100%
    • Average time: 4 sec, 8 sec, 16 sec
    • Pulse rate
      • Measurement range: 20 – 250 bpm
      • Resolution: 1 bpm
      • Accuracy: ±1% or ±1 bpm whichever is greater
    Nellcor SpO2
    • Measurement range: 0-100%
    • Resolution: 1%
    • Accuracy:
      • 70 – 100%: ±2% (adult, pediatric)
      • 70 – 100%: ±3% (neonate)
      • Unspecified (0 – 69%)
    • Alarm range: 0 – 100%
    • Average time: 8 sec, 16 sec
    • Pulse rate
      • Measurement range: 20 – 300 bpm
      • Resolution: 1 bpm
      • Accuracy:
        • 20 – 250 bpm: ±3 bpm
        • 251 – 300 bpm: unspecified
    Masimo SpO2
    • Measurement range: 0 – 100%
    • Resolution: 1%
    • Accuracy:
      • 70 – 100%: ±2% (adult, pediatric, non-motion conditions)
      • 70 – 100%: ±3% (neonate, non-motion conditions)
      • 70 – 100%: ±3% (motion conditions)
      • 0 – 69%: unspecified
    • Average time: 2 – 4 sec, 4 – 6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec
    • Pulse rate
      • Measurement range: 25 – 240 bpm
      • Resolution: 1 bpm
      • Accuracy:
        • ±3 bpm (non-motion conditions)
        • ±5 bpm (motion conditions)
    Select options