Featured Products
- 71,91 лв.Select optionsAC adapter 230/120 V for ri-magic® LED
Specifications
Height 3 cm Weight 50 gm Breadth 4 cm Length 7 cm -
Diagnostic Instruments, Diagnostic Set
415,34 лв.Select optionsRiester e-scope® Oto-/Ophthalmoscope Sets
e-scope® oto-/ophthalmoscopes use advanced optics and LED technology that results in a more efficient diagnosis and reduced maintenance costs. e-scope® oto-/ophthalmoscopes are available with LED, Xenon or vacuum Illumination.
e-scope® otoscope fibre optic version with 2.5 V xenon (3200 K) or innovative 3.7 V LED illumination (5500 K)
- Fibre optics with 2.5 V Xenon (3.200 K) or innovative 3.7 V LED (5.500 K) illumination
- Even field of illumination
- Suitable for pneumatic tests
- 3x magnifying pivoting lens
- Includes reusable ear specula (5 x 2,5mm, 5 x 4mm)
e-scope® otoscope direct illumination with 2.7 V vacuum bulb (2500 K) or 2,5V xenon illumination (3000 K)
- Cost-effective vacuum 2,7V illumination
- Bright 2,5V xenon illumination
- Suitable for pneumatic tests
- 3x magnifying pivoting lens
- including reusable ear specula (5 x 2,5mm, 5 x 4mm)
e-scope® ophthalmoscopes
- With 2.7 V vacuum or 2,5 V xenon illumination and with innovative 3.7 V LED illumination for higher-contrast, brightness and more efficient diagnosis
- Dioptre disc with 18 corrective lenses (+/- 1, 2, 3, 4, 6, 8, 10, 15, 20 dioptres)
- Easy-to-use aperture wheel with six different apertures (fixation star, large circle, small circle, red free filter, blue filter, semi-circle)
- High performance optics with aspherical condenser lens
- Spectacle protection
- Parallel optical path
- Dust-protected
e-scope® otoscopes and ophthalmoscopes are available in an impact-resistant case or in a space-saving bag including accessories (ear specula).
Specifications
Height 5 cm Weight 350 gm Breadth 10 cm Length 15 cm - From 4801,04 лв.Select options
Riester RVS 100 Advanced Vital Signs Monitor
Advanced vital signs monitor for pulse oximetry, blood pressure and temperature measurements.
Riester’s RVS-100 is an advanced vital signs monitor which offers pulse oximetry, blood pressure and temperature measurements. The device has ‘Monitor‘, ‘Spot-Check‘ and ‘Triage‘ profile options and a high performance touchscreen.
The RVS – 100 can communicate through a wired or wireless connection to the hospital’s EMR, according to the HL7 standard. The highly flexible, modular design offers many configurations and settings to suit the needs of hospitals, clinics and private practices, as well as long-term care settings.
- 3 measurement modes (Monitor, Spot-check, Triage)
- High-sensitivity 8” TFT colour touch screen with self-explanatory user interface and integrated help- function
- Displays SYS, DIA, MAP, Pulse, Temp, Plethysmogram
- Clinical grade NIBP* including averaging mode
- Programmable NIBP averaging mode (2-5 readings)
- Riester, Masimo* or Nellcor* Oximax SpO2
- Predictive Thermometer (Filac 3000)
- Connectivity with the EMR – follows the HL7 standard
- LAN Connectivity as standard
- Programmable alarms (audible, visible) with 3 priority levels
- Supports nurse call RJ11 port
- Available as wall, desk and floor model
- Can be expanded to wall diagnostic station
- Internal memory for 5,000 measurements
- Lithium-ion battery with up to 17 hours ON time
Options
- Thermal printer (internal)
- Barcode reader
- Mobile stand including basket for storage (cuffs, sensors, cables)
- WiFi
Accessories supplied with monitor
- NIBP extension tube
- 2 cuffs (adult & adult large incl. connectors)
- SpO2 sensor adult
- SpO2 extension cable
- Oral temperature probe
- Probe cover
Technical Data
Monitor
- Power Supply: 100 – 240 V AC, 50/60 Hz
- Battery: Lithium-Ion battery, rechargeable, 10.8 V DC, 6,600 mAh,
Charging time to a 100%: 6 hours,
- Operating time approx. 11 hours (new and fully charged battery at 25°C ambient temperature with connected SpO2, Temp, and NIBP on AUTO mode (15 minutes interval))
- Low battery indicator
- Charging battery indicator
- Dimensions Monitor: L x H x W: 303 x 230 x 110 mm (11.93 x 9.06 x 4.33 in)
- Weight: 3.60 kg (7.94 lb) including temp module, printer
- Dimensions Basket: L x H x W: 270 x 225 x 200 mm (10.63 x 8.86 x 7.88 in)
Alarm:
- Visible LEDs (Yellow, Red, Blue, Green, Orange)
- Audible Speaker
- Gives audible alarm, QRS tone
- Supports Pitch Tone and multi-level volume
- Alarm tones meet the requirement of IEC 60601-1-8.
- Alarm Pressure: 45 dB to 85 dB test distance is 1 meter from the tone.
Ports:
- 1 x AC power inlet
- 1 x standard RJ45 interfaces.100 BASE-TX, IEEE 802.3
- 4 x USB ports
- 1 x Equipotential grounding point
- 1 x RJ11 connector for nurse call
- DC out port 15V/1.2 A Clinical-grade NIBP (non-invasive blood pressure)
Clinical-grade NIBP (non-invasive blood pressure)
- Measuring method: automatic oscillometric
- Measurement range:
- Adult SYS 30 – 270 mmHg, DIA 10 – 220 mmHg, MAP 20 – 235 mmHg
- Pediatric SYS 30 – 235 mmHg, DIA 10 – 220 mmHg, MAP 20 – 225 mmHg
- Neonate SYS 30-135 mmHg, DIA 10-110 mmHg, MAP 20-125 mmHg
- Cuff pressure range: 0 – 280 mmHg
- Resolution: 1 mmHg
- Pressure accuracy:
- Static ±3 mmHg
- Clinic ±5 mmHg, standard deviation ≤ 8 mmHg
- Pulse rate range: 40 – 240 bpm
- Pulse rate accuracy: 1 bpm
- Initial inflation pressure:
- Adult 160 mmHg, Pediatric 130 mmHg, Neonate 75 mmHg
- Double overpressure protection (hardware and software)
- Adult 297 ± 3 mmHg, Pediatric 252 ± 3 mmHg , Neonate 147 ± 3 mmHg
- Alarm range:
- SYS : 0 – 300 mmHg
- DIA : 0 – 300 mmHg
- MAP : 0 – 300 mmHg
Riester SpO2
- Measurement range: 0 – 100%
- Resolution: 1%
- Accuracy:
- ±2% (70 to 100%)
- ±3% (40% to 69%)
- Unspecified (0% to 39%)
- Alarm range: 0 – 100%
- Average time: 4 sec, 8 sec, 16 sec
- Pulse rate
- Measurement range: 20 – 250 bpm
- Resolution: 1 bpm
- Accuracy: ±1% or ±1 bpm whichever is greater
Nellcor SpO2
- Measurement range: 0-100%
- Resolution: 1%
- Accuracy:
- 70 – 100%: ±2% (adult, pediatric)
- 70 – 100%: ±3% (neonate)
- Unspecified (0 – 69%)
- Alarm range: 0 – 100%
- Average time: 8 sec, 16 sec
- Pulse rate
- Measurement range: 20 – 300 bpm
- Resolution: 1 bpm
- Accuracy: 20 – 250 bpm: ±3 bpm & 251 – 300 bpm: unspecified
Masimo SpO2
- Measurement range: 0 – 100%
- Resolution: 1%
- Accuracy:
- 70 – 100%: ±2% (adult, pediatric, non-motion conditions)
- 70 – 100%: ±3% (neonate, non-motion conditions)
- 70 – 100%: ±3% (motion conditions)
- 0 – 69%: unspecified
- Average time: 2 – 4 sec, 4 – 6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec
- Pulse rate
- Measurement range: 25 – 240 bpm
- Resolution: 1 bpm
- Accuracy: ±3 bpm (non-motion conditions) & ±5 bpm (motion conditions)
Predictive thermometer
- Oral ( Quick Mode )
- non-fever temps: 3-5 seconds
- fever temps: 8-10 seconds
- Oral ( Standard Mode ): 6-10 seconds
- Axillary Mode: 8-12 seconds
- Rectal Mode: 10-14 seconds
- Direct Mode (all sites): 60-120 seconds
- Measurement accuracy: 35.5ºC – 42ºC (95.9°F – 107.6°F)
- Direct Mode (all sites): ±0.1ºC (±0.2ºF)
- Standard Prediction Mode (all sites): ±0.1ºC (±0.2ºF)
- Quick Prediction Mode (oral only): ±0.3ºC (±0.5ºF)